INDICATIONS
VANTAS is indicated for the
palliative treatment of advanced prostate cancer.
DOSAGE AND ADMINISTRATION
Recommended Dose
The recommended dose of VANTAS
is one implant for 12 months. Each implant contains 50 mg histrelin acetate to
deliver 41 mg histrelin. The implant is inserted subcutaneously in the inner
aspect of the upper arm and provides continuous release of histrelin (50
mcg/day) for 12 months of hormonal therapy. VANTAS should be removed after 12
months of therapy (the implant has been designed to allow for a few additional
weeks of histrelin release, in order to allow flexibility of medical
appointments). At the time an implant is removed, another implant may be
inserted to continue therapy.
Recommended Procedure For Implant
Insertion And Removal
This procedure section is
intended to provide guidance for the insertion and removal of VANTAS.
Insertion of a new implant can
proceed using the following Insertion Procedure. If a previous VANTAS
implant must first be removed, please see the Removal Procedure instructions
below.
Insertion Procedure
Many of the supplies necessary
to insert the implant, including the Insertion Tool and local anesthetic, are
provided in a separate Implantation Kit that is shipped along with the implant.
Please note that the implant, in its sealed vial, pouch, and carton, must be
kept refrigerated (2-8°C) until needed for the procedure. Once removed from
refrigeration, the vial containing the implant (still in its unopened pouch and
carton) may remain at room temperature for up to 7 days, if necessary, before
being used. If not used in that time, the packaged implant may again be properly
refrigerated until the expiration date on the carton.
NOTE: The Implantation Kit is
to be stored at 20° to 25°C (68° to 77°F) [See USP Controlled Room
Temperature] only.
Insertion of the VANTAS implant
is a surgical procedure. Sterile gloves and aseptic technique must be used to
minimize any chance of infection.
Setting Up The Sterile Field
Using proper aseptic technique, the sterilized components
of the Implantation Kit needed for the insertion procedure, including the
Insertion Tool, are to be carefully dispensed from their packaging onto the
Sterile Field drape (nonfenestrated) provided. NOTE THAT THE KIT BOX AND ALL
PACKAGING, INCLUDING THE SURFACE OF THE VIALS, ARE NOT STERILE and should be
kept off of the Sterile Field drape.
The implant vial should not be opened until just before
the time of insertion. Open the vial by removing the metal band and carefully
pour the sterile contents (implant and sterile saline) onto the Sterile Field
drape. The implant can then be handled with sterile gloves or with the sterile
mosquito clamp provided. AVOID bending or pinching the implant.
Preparing The Patient And The Insertion Site
The patient should be on his back, ideally with the arm
least used (e.g., left arm for a right-handed person) positioned either bent or
extended so that the physician has ready access to the inner aspect of the
upper arm. Propping the arm with pillows may help the patient more easily hold
the position. The suggested optimum site for subcutaneous insertion is approximately
half-way between the shoulder and the elbow, in line with the crease between
the biceps and triceps muscles.
Antiseptic
Swab the insertion area with
topical antiseptic, then overlay with the fenestrated Sterile Field
drape provided, so that the opening is over the insertion site (for clarity of
illustration, the following images do not show the drape).
Anesthetic
The method of anesthesia utilized (i.e., local, conscious
sedation, general) is at the discretion of the healthcare provider.
If local anesthesia is selected:
a vial of a local anesthetic (e.g., lidocaine HCl 1% with epinephrine or
lidocaine HCl 1%) (note that the exterior of the vial is not sterile) has been
provided along with a sterile hypodermic needle for injection. After
determining the absence of known allergies to the anesthetic agent, inject
anesthetic into the subcutaneous tissue, starting at the planned incision site,
then infiltrating along the intended subcutaneous insertion path, up to the
length of the implant (a little more than one inch).
Loading The Insertion Tool
The sterile Insertion Tool is
comprised of a fixed handle attached to a retractable, bevel-tipped cannula,
into the chamber of which the implant is placed for subcutaneous insertion.
Inside the fully extended cannula, up to the level of the black marking, is a
fixed piston upon which the implant rests. During the final step of the
insertion procedure, the cannula will be retracted into the handle using the
slide mechanism (green button), thereby exposing and leaving the implant to remain
in the subcutaneous tissue.
When first grasping the sterile
Insertion Tool, confirm that the cannula is fully extended. Verify this by
inspecting the position of the green retraction button. The button should be
locked in position all the way forward, towards the cannula, farthest from the
handle.
The implant can be picked up
using sterile gloves or with the sterile mosquito clamp provided. Avoid bending
or pinching the implant. Note that the implant may come out of its vial
slightly curved and/or partially flattened after refrigerated storage.
To help make the implant more
symmetrical prior to loading into the Tool, you can roll the implant a few
times (using a sterile glove) between the fingers and thumb.
Insert the implant into the
cannula of the Insertion Tool manually or using the mosquito clamp. When
inserting the implant into the cannula, DO NOT FORCE the implant. If resistance
is felt, the implant should be removed and manually manipulated or rolled as
needed, and re-inserted into the cannula.
or
When fully inserted, the
implant rests inside the cannula so that just the tip of the implant is visible
at the beveled end of the cannula.
Making The Incision
Using the sterile scalpel provided, make an incision
transverse to the long axis of the arm, and of a size adequate to allow the
bore of the cannula to be inserted into the subcutaneous tissue.
Inserting The Implant
The insertion may be easier if a “pocket” for the implant
is first created by blunt dissection through the incision, subcutaneously along
the path of the anesthetic, using the cannula of the loaded Insertion Tool, or
using a sterile hemostatic clamp or equivalent surgical tool.
Be sure to VISIBLY RAISE THE
SKIN (known as tenting) at all times during the pocket-making and insertion
procedures to ensure correct subcutaneous placement (“just under the skin”) of
the implant. Note that the cannula of the Insertion Tool, or whatever tool is
being used to create the pocket, SHOULD NOT ENTER MUSCLE TISSUE. Deep insertion
of the implant will not affect the performance of VANTAS, but may cause
difficulty in the later removal of the implant.
If using the cannula of the
loaded Insertion Tool to create the pocket, carefully insert the tip of the
cannula into the incision and advance through the subcutaneous tissue, while
visibly raising the skin along the length of the cannula up to, but no farther
than, the inscribed black line on the cannula. DO NOT DEPRESS THE GREEN
RETRACTION BUTTON ON THE TOOL WHILE INSERTING OR ADVANCING THE TOOL INTO THE
INCISION.
Pull the Tool back, almost to
the beveled tip of the cannula, and advance the Tool forward again, so that the
cannula reenters the pocket completely, but no farther than the inscribed black
line. Be sure to keep the insertion path just immediately subcutaneous.
If another tool was used to
create the pocket, now insert the loaded cannula of the Insertion Tool
containing the implant through the incision, up to the inscribed black line.
Hold the Insertion Tool in
place with the base against the patient's arm as you carefully move your thumb
to the green retraction button. Depress the button to release the locking
mechanism, then slide the button back toward the handle until it stops, all the
while holding the body of the Insertion Tool in place.
Retracting the button causes
the cannula to withdraw from the incision, leaving the implant in the
subcutaneous tissue. DO NOT FURTHER ADVANCE THE CANNULA ONCE THE RETRACTION
PROCESS HAS STARTED. Likewise, do not withdraw the Insertion Tool until the
button is fully retracted or the implant may be pulled partially out of the
incision. Once the retraction is complete, the Tool can be fully withdrawn.
NOTE: It is helpful during the process of retraction and
withdrawal of the cannula to apply pressure to the skin over the implant, to
help ensure that the implant remains in the subcutaneous pocket.
If there is a need to re-start
the process at any time during the insertion procedure, withdraw the Insertion
Tool, carefully extract the implant from the cannula and reset the retraction
button on the Tool to its forward-most position. Examine the implant before
reloading the implant into the Insertion Tool, and start again.
Placement of the implant should be confirmed by
palpation. Note that the tip of a properly-placed implant may not be visible
through the incision.
After implantation, briefly cover the site with a sterile
gauze pad and apply pressure to ensure hemostasis.
Closing The Incision
To close the incision, you can use the absorbable sutures
and/or the sterile adhesive surgical strips provided. To improve adhesion of
the strips, you can apply benzoin tincture antiseptic (provided) to the skin,
and let it dry, before applying the adhesive strips.
Once closed, cover the incision
site with sterile gauze pads and secure the dressing with the bandage provided.
Please provide the patient with
a Patient Information Leaflet, which includes information about the implant and
instructions on proper care of the insertion site.
Removal Procedure
VANTAS should be removed after
12 months of therapy. Most of the supplies necessary to remove the implant,
including the local anesthetic and the sterile mosquito clamp, are provided in
the Implantation Kit that is shipped along with a new VANTAS implant. Note that
the Implantation Kit is to be stored at 20° to 25°C (68° to 77°F) [See USP
Controlled Room Temperature] only. See the Insertion Procedure above
for further instructions.
Removal of the VANTAS implant
is a surgical procedure. Sterile gloves and aseptic technique must be used to
minimize any chance of infection.
Setting Up The Sterile Field
Using proper aseptic technique, the sterilized components
of the Implantation Kit needed for the implant removal procedure are to be
carefully dispensed from their packaging onto the Sterile Field drape (non-fenestrated)
provided. NOTE THAT THE KIT BOX AND ALL PACKAGING, INCLUDING THE SURFACE OF THE
VIALS, ARE NOT STERILE and should be kept off of the Sterile Field drape.
Preparing The Patient And The Site
The patient should be on his back, with the arm
containing the implant positioned, either bent or extended, so that the
physician has ready access to the inner aspect of the upper arm. Propping the
arm with pillows may help the patient more easily hold the position.
The implant to be removed should first be located by
palpating the inner aspect of the upper arm, near the incision from the prior
year.
Generally, the previous implant
is readily palpated. In the event the implant is difficult to locate,
ultrasound may be used. If ultrasound fails to locate the implant, other
imaging techniques such as CT or MRI may be used to locate it (plain films are not
recommended as the implant is not radiopaque).
Antiseptic
Swab the area above and around
the previous implant with topical antiseptic. Overlay the area with the fenestrated
Sterile Field drape provided, so that the hole is over the previous insertion
site (for clarity of illustration, the following images do not show the drape).
Anesthetic
The method of anesthesia
utilized (i.e., local, conscious sedation, general) is at the discretion of the
healthcare provider.
If local anesthesia is selected:
a vial of local anesthetic (e.g., lidocaine HCl 1% with epinephrine or
lidocaine HCl 1%) (note that the exterior of the vial is not sterile) has been
provided along with a sterile hypodermic needle for injection. After
determining the absence of known allergies to the anesthetic agent, inject
anesthetic into the subcutaneous tissue at and around the site of the intended
incision (the site of the previous implant).
Making The Incision And Removing
The Implant
Using the sterile scalpel
provided, make an incision of a size adequate to allow the implant to be easily
removed and, if a new implant will be inserted, large enough for the bore of
the cannula of the Insertion Tool provided.
Generally, the tip of the
implant will be visible through the incision, possibly covered by a
pseudocapsule of tissue. In order to facilitate the removal of the implant, it
may be necessary to palpate the head of the implant through the incision using
your smallest finger, especially if the head of the implant is not readily
visible. In addition, you may need to push down on the distal end of the
implant and “massage it forward” toward the incision.
Carefully nick the
pseudocapsule to reveal the polymer tip of the implant. It may be beneficial to
insert the sterile mosquito clamp provided into the hole created in the
pseudocapsule and expand by opening the clamp. Widening the opening of the
pseudocapsule may ease the extraction of the implant.
Gently but securely grasp the
implant with the sterile mosquito clamp and extract the implant.
Dispose of the implant in a proper manner, treating it
like any other biowaste.
Briefly cover the site with a sterile gauze pad and apply
pressure to ensure hemostasis.
If inserting a new implant, see the Insertion
Procedure instructions provided above. Note that you can insert the new
implant into the same “pocket” as the removed implant, or make a new incision
at a different site in the same arm or in the contralateral arm.
If a new implant is not to be inserted, proceed to close
the incision.
Closing The Incision
To close the incision, you can use the absorbable sutures
and/or the sterile adhesive surgical strips provided. To improve adhesion of
the strips, you can apply benzoin tincture antiseptic (provided) to the skin,
and let it dry, before applying the adhesive strips.
Once closed, cover the incision
site with sterile gauze pads and secure the dressing with the bandage provided.
HOW SUPPLIED
Dosage Forms And Strengths
VANTAS® (histrelin acetate) implant is a sterile,
non-biodegradable, diffusion-controlled hydrogel polymer reservoir containing
histrelin acetate, a synthetic nonapeptide analog of the naturally occurring
gonadotropin releasing hormone (GnRH). VANTAS is designed to deliver
approximately 50 mcg histrelin acetate per day (equivalent to approximately 41
mcg histrelin per day) over 12 months.
Storage And Handling
VANTAS (NDC 67979-500-01) is supplied in a carton
containing 2 inner cartons, one for the VANTAS implant and one for the VANTAS
Implantation Kit:
The VANTAS implant contains 50 mg of histrelin acetate.
The VANTAS implant carton contains a cold pack for refrigerated shipment and a
small carton containing an amber plastic pouch. Inside the pouch is a glass
vial with a Teflon-coated stopper and an aluminum seal, containing the implant
in 2 mL of sterile 1.8% sodium chloride solution. (Note: The 3.5 mL vial is not
completely filled with saline.)
Upon receipt, refrigerate the small carton containing
the amber plastic pouch and glass vial (with the implant inside) until the day
of insertion. The implant vial should not be opened until just before the time
of insertion.
Store the implant refrigerated, 2-8°C (36-46°F), in the
unopened glass vial with the sterile 1.8% sodium chloride solution, overwrapped
in the amber plastic pouch and carton, until the expiration date provided.
Excursion permitted to 25°C (77°F) for 7 days. Protect from light. Do not
freeze.
Store the VANTAS Implantation Kit at room temperature
only.
The VANTAS Implantation Kit carton contains one each of
the following (individually wrapped in sterile packaging): implant insertion tool,
#15 disposable scalpel, syringe with 18 gauge needle, 25 gauge 1.5” needle, SS
mosquito clamp, benzoin tincture antiseptic, alcohol swabs (2 packages),
fenestrated drape, non-fenestrated drape, skin antiseptic swab, gauze sponges,
surgical closure strips, coated absorbable sutures, cohesive bandage, and a local
anesthetic (e.g., lidocaine HCl 1% with epinephrine or lidocaine HCl 1%).
Manufactured by: Endo Pharmaceuticals Solutions Inc.
Malvern, PA 19355. Revised: Feb 2019